General Instruments Precision Industrial RTD’s
GIC is the leading Precision Industrial RTD’s in India providing reliable solutions for accurate temperature measurement to process industry worldwide.
Any industrial installation has several electrical thermometers such as Thermocouples & RTDs. At a regular interval, these instruments are required to be checked against a ‘Standard Industrial RTD’, which has accuracy to the level of 1/10th DIN. General with its vast experience in design and manufacturing of precision instruments has developed ‘Standard PT100 Industrial RTD’ which caters to such requirements of client. These RTDs are traceable to National Physical Laboratory (NPL) and are either provided with Original Calibration Certificate from NPL with very low level of uncertainty or Internal Certificate traceable to National laboratory. These Standard RTDs are manufactured with 4 wire design to deliver highest performance.
Routine tests:
a) Accuracy/Calibration Test
b) Nitrogen Leakage Test (Weld
Closure Conformity)
c) Insulation Resistance Test
d) Thermal Response Time Test
e) Continuity Test
f) Dimension Checks
Type tests:
a) Vibration test
b) Drop / Shock test
c) Self-heating error test
d) Hysteresis Test
e) Stability Test
f) Autoclave Test
g) Pull Test (Load Withstand Test)
h) Di-electric strength Test (H.V. Test)
Features:
- Resistance : Nominal 100 ohms at 0°C (PT100) Four Wire System.
- Temperature Coefficient : 0.00385 ohms/ohm/°C nominal
- Temperature Range : (–)200°C to 500°C (transition and cable temperature 150°C maximum)
- Resistance Stability : ±0.20 °C at 0°C after 1000 hours at 400°C
- Sheath Material : Inconel 600
- Leads : Teflon™-insulated, silver-plated stranded copper, 22 AWG. Colour Coded as 2 White / 2 Red
- Termination : Four gold-plated spade lugs are standard. Other options available.
- Hysteresis : 0.04% maximum between (–)200°C and 500°C
- Time Constant : Four seconds maximum for 63.2% response to step change in water moving at 1mps.
- Conductor : Pure Silver
- Calibration : Includes National Physical Laboratory Certificate in Original. (Calibration points (-)25°C, 0°C, 100°C, 200°C, 300°C, 400°C & 500°C) Any other points optional.
- Probe Accuracy : Includes calibration uncertainty and short-term stability (-)25°C ±0.070°C, 0°C ±0.030°C, 200°C ±0.10°C, 300°C ±0.15°C
- Immersion Effects : Reading will not vary more than 0.02°C when the probe immersion is varied between 50 mm and 150 mm in an ice bath.
- Size : PIR01-300: 300 mm L C 6.0 mm Dia. PIR01-500: 500 mm L C 6.0 mm Dia.